Troubleshooting: How do I limit ground loops?

问题:

Troubleshooting: How do I limit ground loops?

答案:

The preferred method for reducing ground loops is to ground the measurement circuit in only one place. The most common problem is that Measurement LO input of the meter is connected to ground at the instrument, and ground is connected to the circuit under test a few feet away. Even when the grounding points are in close proximity, the point to which the ground is attached matters greatly.

此常见问题适用于:

产品系列: Keithley 2182A 型纳伏表

产品:

常见问题 ID 70556