gongjigongyi.com

与米乐m6网页版登录入口代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

How do I configure the LAN settings on an MDO4000C series oscilloscope?

问题:

How do I configure the LAN settings on an MDO4000C series oscilloscope?

答案:

This procedure also applies to the DPO/MSO/MDO4000B series oscilloscopes. 

As it stands now, all 4000 series scopes  require a live network connection  be made on the scope in order to set and save LAN settings. If the scope does not have a network connection, the IP address will default to 0.0.0.0. This can be seen by going to the Utility menu, scrolling to the I/O page under the Config Menu, Selecting Network Configuration, selecting manual, and selecting Set IP Addresses Manually. Set any IP address you like. If you go back to the Ethernet & LXI settings menu, and select LAN settings, the IP address is 0.0.0.0 because there is no network connection. But as soon as you plug an Ethernet cable connected to a network into the back of the scope, the scope sets the IP address that you specified manually.
 
The same behavior occurs when using Automatic network configuration without a network connection. Remove the network connection by unplugging the Ethernet cable from the back of the scope. Go to Network Configuration and select Automatic (DHCP & Auto-IP). Go back to Ethernet & LXI and LAN Settings and you’ll see that the IP address has reverted back to 0.0.0.0. But again, as soon as you reenable the network connection, the scope automatically obtains the IP addresses it needs.  


此常见问题适用于:

产品系列: MSO4000 / DPO4000 MDO4000C混合域示波器

产品:

常见问题 ID 66236

查看所有常见问题 »