gongjigongyi.com

与米乐m6网页版登录入口代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

ExceLINX: Why do I get no measurements when using the KPCMCIA-GPIB (Ines) Card with ExceLINX-1A?

问题:

ExceLINX: Why do I get no measurements when using the KPCMCIA-GPIB (Ines) Card with ExceLINX-1A?

答案:

Why do I get no measurements when using the KPCMCIA-GPIB (Ines) Card with ExceLINX-1A?

There is a known problem with ExceLINX-1A (ver C01) software for Integra series 2700 or 2750 when using the DMM Scan task when the GPIB interface is the Ines or KPCMCIA-GPIB.

The DMM Config sheet will properly execute, but the DMM Scan task will show the Status "RUNNING" but it never really triggers the 27xx to make a measurement. Therefore no readings are returned.

Workaround: in the DMM Config sheet, set the FRONT PANEL LOCK OUT setting to "OFF". The default setting is ON. If the setting is left to the default, then you will see the above symptoms. The Front Panel Lock out enables or disables the Front Panel buttons of the 27xx.

Root Cause: In the Ines GPIB driver, the high level GoToLocal function used in the passport does not return. Use of the low level GTL bus command executes properly.

Correction: a corrected passport for Ines GPIB cards will be distributed in the next maintenance release of ExceLINX-1A (ver C02).

 


此常见问题适用于:

产品系列: Keithley 2700 万用表/数据采集/开关系统

产品:

常见问题 ID 70156

查看所有常见问题 »